Detection target
Determination of ethylene oxide content
Introduction
Ethylene oxide (EO), as a highly effective low-temperature sterilizing agent, is widely used in the sterilization of medical devices (especially those made of heat-sensitive plastics, rubber, and polymer materials, such as face masks). However, residual ethylene oxide and its reaction products (such as 2-chloroethanol) have potential carcinogenic, mutagenic, and irritant effects, which may pose risks to the health of patients and healthcare workers. Therefore, the residual ethylene oxide levels in medical devices must be strictly tested and controlled before they are placed on the market to ensure that the biological safety of the products meets relevant international standards (such as ISO 10993-7) and various regulatory requirements. Gas chromatography is typically used to quantitatively assess the residual ethylene oxide in products after a specified aeration period following sterilization. This analysis is a critical step in the quality control and registration of medical devices.
Overview
This solution conforms to the ISO 10993.7 Medical devices. Biological evaluation of medical devices. Part 7: Ethylene oxide sterilization residuals. The instrument tests the residual amount of ethylene oxide in medical devices after ethylene oxide sterilization. The standard requires the detection of extremely low levels of residue (EO limits are typically measured in μg/g), therefore this instrument offers the advantages of high sensitivity and a low detection limit.
Apparatus
HKL-10993 Gas Chromatograph for Ethylene Oxide

Configuration
Serial | Name | Specification | Number |
1 | Gas Chromatograph(HM-GC) Ethylene Oxide Specialized | 1) 5.7 LCD display,Startup self-check,Wide-range self-diagnosis, Intelligent fuzzy control rear-door system, Over-temperature protection,7-step temperature programming 2) Configuration: Main unit + FID + capillary inlet | 1 set |
2 | Headspace Sampler | Semi-automatic | 1 set |
3 | Ethylene Oxide Specialized Column | Capillary column 30m*0.32μm*0.5mm | 1 pc |
4 | Ethylene Oxide Standard Sample | For calibration curve | 1 large bottle |
5 | Workstation | Data acquisition | 1 set |
6 | Air Generator | 1) Air purity: Oil-free, triple filtration 2) Output flow: 0-3 L/min 3) Input pressure: 0-0.4MPa (adjustable) 4) Rate power: 150W | 1 set |
7 | Hydrogen generator | 1) Hydrogen purity: 99.999% 2) Output flow: 0-300 ml/min 3) Input pressure: 0-0.4 Mpa 4) Rate power: <150W | 1 set |
8 | Nitrogen cylinder | Purity: 99.999%; includes pressure regulator | User-provided |
9 | Computer + printer | User-provided |